Calibration

 Calibration data are taken obtained using Special Requirements. The use of the APT Parameter Field CALIBRATION Special Requirement is described below.

Calibration Special Requirement

APT Restricted Field CALIBRATION APT Value Field CALIBRATION TYPE APT Value Field LAMP STATE

The APT Parameter Field CALIBRATION special requirement is motivated by the need to enable certain modes of observation that are considered only desirable for calibration observations.  Some such modes activate internal lamps or block outside light, and as such can be detrimental to general astrophysical observations.  As a consequence, the APT Parameter Field CALIBRATION special requirement is only available when APT Value Field Allow Restricted is selected (see Program Information).

The APT Parameter Field CALIBRATION special requirement is applied individually to selected observation specifications within a Pass Plan

This requirement specifies that the selected observation specification is used for calibration, using the specified APT Value Field CALIBRATION TYPE and APT Value Field LAMP STATE .  The APT Value Field CALIBRATION TYPE chosen determines whether the APT Value Field DARK element is selected in the Element Wheel, thus blocking external light, and whether the internal light source is turned on with the appropriate specifications indicated in APT Value Field LAMP STATE .  Details are given in Table 1 below.

APT Restricted Field CALIBRATION   and APT Restricted Field ENGINEERING special requirements are not allowed on the same observation.

<CALIBRATION TYPE>

Calibration types are: Dark Imaging, Dark Spectroscopy, Internal Flat, CRNL Direct Illumination, CRNL LOLO. APT Value Field CALIBRATION TYPE is required.

Depending on the APT Value Field CALIBRATION TYPE , certain parameters of the observation specification are set to required values and cannot be changed (see Table 1). 

Calibration Types Dark Imaging and Dark Spectroscopy are used to obtain dark images to generate dark calibration reference frames.  These images will be taken with the APT Value Field DARK element in the Element Wheel and with lamps turned on.  No dither shall be specified, although a repeat count is allowed.  These observations will typically use the diagnostic MultiAccum table and use guide windows of sizes appropriate to Imaging (16x16) and Spectroscopy (32x16), respectively.  The position of the guide window needs to be randomized.

Calibration types Internal Flat and CRNL Direct Illumination are used to obtain frames illuminated by the internal light sources.  The APT Value Field DARK element of the Element Wheel is in place.   APT Value Field LAMP STATE specifies how many exposures are taken and at which lamp wavelength and illumination level.  Typically, Internal Flats will be obtained with a bright lamp setting suitable to collect about 50k electrons per pixel in a 50s exposure.  CRNL Direct Illumination data will be obtained with a combination of two lamps at the same wavelength, turned on at different settings; a sequence of several dozen such combinations will be taken in order to calibrate the nonlinearity of the detector using the Combinatorial Flux Addition method.  More information under APT Value Field LAMP STATE .

Calibration Type CRNL LOLO is used to obtain an image of the sky superimposed with light from the internal source.  Pointing, APT Parameter Field OPTICAL ELEMENT , and APT Parameter Field DITHER are defined as with any other exposure.  A lamp associated with the selected optical element needs to be turned on at the desired intensity; lamp light is back-reflected by the cold mask on the filter and added to the scene light.   APT Warning Note that there is no lamp associated with the optical element F213.

Table 1. Required Observation Specification parameters by APT Value Field CALIBRATION TYPE

APT Value Field CALIBRATION TYPE

APT Parameter Field TARGET

APT Parameter Field OPTICAL ELEMENT

APT Parameter Field MOSAIC

APT Parameter Field DITHER

APT Parameter Field SUBPIXEL DITHER

APT Parameter Field EXPOSURES/DITHER

Dark ImagingNONE

APT Value Field DARK

APT Value Field NONE

APT Value Field NONE

APT Value Field NONE


Dark SpectroscopyNONE

APT Value Field DARK

APT Value Field NONE

APT Value Field NONE

APT Value Field NONE


Internal FlatNONE

APT Value Field DARK

APT Value Field NONE

APT Value Field NONE

APT Value Field NONE

Equal to number of exposures in the lamp state
CRNL Direct IlluminationNONE

APT Value Field DARK

APT Value Field NONE

APT Value Field NONE

APT Value Field NONE

Equal to number of exposures in the lamp state
CRNL LOLO(selected fixed target)

(selected optical element)

(selected mosaic)

(selected dither)(selected subpixel dither)Equal to number of exposures in the lamp state

<LAMP STATE>

APT Value Field LAMP STATE specifies the state of active lamps for each exposure. 

There are 12 LED lamps (2 pairs at each of 6 different wavelengths). Calibrations using the lamps may have sequences up to 100 exposures with different subsets of lamps selected for each exposure and different power levels configured for each lamp. Lamp power levels (lamp states) are discrete integers ranging from 1 to 128.

APT Value Field LAMP STATE is available and required when APT Value Field CALIBRATION TYPE is Internal Flat, CRNL Direct Illumination, or CRNL LOLO.  The use of the lamp is different for each of these cases.

For Internal Flat observations, one lamp is turned on for one of the six bands at a high flux level, suitable to collect 50K electrons per pixel per exposure.  (A typical exposure may be around 50s.)  Multiple exposures are obtained until at least 10^6 electrons per pixel are collected in that band.  The process is then repeated for the other 5 bands.  Two lamps for the same band can be turned on if needed to achieve the desired luminosity.  The APT Value Field DARK element is in use throughout, and the light is diffused by the back of the APT Value Field DARK element.

For CRNL Direct Illumination observations, a sequence of observations is taken with both lamps in the same band turned on.  During the sequence, the intensity setting of both lamps is cycled through several values, resulting in images with multiple levels of illumination.  The Combinatorial Flux Addition method is then used to determine any deviation from linearity in the detector response (after correction for classic nonlinearity).  This procedure can constrain non-linearity over a broad range of impinging fluxes and total count levels.  The procedure is then repeated for other bands as needed.  As for the Internal Flat observations, the lamp light is diffused by the back of the Dark element, which is in place throughout.  Note that the exposure length may need to change depending on the lamp setting; at low intensity, long exposures are needed to collect enough counts, while at high intensity, short exposures are needed to avoid saturation.  Using the number of elements in the sequence to determine number of exposures is only useful for sequences of exposures reaching similar level of total lamp flux.

For CRNL LOLO, the telescope is pointed at a carefully chosen external scene, with the desired filter, pointing, and dithering options.  The CRNL LOLO observations are characterized by turning on one lamp corresponding to the optical element chosen; its light is diffused by the cold mask in the filter, and added to the signal received by the detector.  This provides the essence of the Lamp-On/Lamp-Off (LOLO) process; the change in signal in a source when a smooth illumination (from the internal lamp) is added provides constraints on the detector linearity at that signal level.  Typically we expect the observing sequence to be repeated with lamp off, lamp on at low setting, and lamp on at higher setting. APT Warning Note that there is no lamp suitable for use with F213.

<LAMP STATE> entry format

The format of APT Value Field LAMP STATE is multiple lines, each starting with an exposure number followed by some number of lamp state directives. Each element, including the exposure number, must be separated by a comma.

A APT Value Field LAMP STATE directive is an 'L' followed by the lamp number, then an '=', then the lamp state as an integer in 1-128. Ex: L10 = 22.
Whitespace around the '=' is optional but should not be present in the exported SQL.

An example:

1, L10 = 33, L2 = 20
2, L3 = 40, L5 = 21

 <LAMP STATE> Usage

Table 2. Lamp usage where APT Value Field EXPOSURE NUMBER , L APT Value Field LAMP NUMBER = APT Value Field LAMP STATE

Element

Lamp Usage

APT Value Field EXPOSURE NUMBER

sequential number from APT Range Field 1-100 , must start with 1

APT Value Field LAMP NUMBER

lamp number from APT Range Field 1-12

APT Value Field LAMP STATE

lamp state number from APT Range Field 1-128 (power level)


The image is a screen capture of APT showing the various fields available in APT when using the lamps to observe an internal flat field.


Figure 1: APT Value Field LAMP STATE entry in the Roman APT user interface. The entry above would generate 20 internal flat exposures with lamp 12 at power level 5.



For additional questions not answered in this article, please contact the Roman Help Desk at STScI.



Latest Update

Publication

 

Initial publication of the article.